Continuous advances in high-speed communication and measurement systems require higher levels of
performance from system clocks and references. Performance acceptable in the past may not be sufficient
to support high-speed synchronous equipment. Perhaps the most important and least understood measure
of clock performance is jitter.
The purpose of this discussion is fourfold.
The expected edges in a digital datastream never occur exactly where desired. Defining and measuring
the timing accuracy of those edges (jitter) is critical to the performance of synchronous communication
systems.
Digital Waveform with Jittered Edges
Figure 1
Definition of terms:
For example:
Assume a clock rate of 155.52 MHz. One unit interval would be equal to the period of the signal, 1/155.52
MHz = 6.43 nsec. = 360 deg.
Assume 100 ps Pk-Pk of jitter.
100 Ps of jitter = .01555 unit intervals (UI) of jitter = 5.598 deg. of jitter. (All Pk-Pk)
All three measurements describe the same amount of jitter.
For jitter power, rms (one sigma,
) measurements are used. For the above
case, we approximate Pk-Pk as 7
, or 7 times the RMS value, placing the rms.
jitter power at .0000049 UI2. {(.01555/7)2 . Expressed in dB, relative to one
unit interval, jitter power in this case would be 10log (.0000049)= -53.1 dB ui. As will be
seen later, jitter can be derived from power spectral density (phase noise) measurements. Table 1
relates various measures of jitter in a 155.52 MHz system clock.
| Pk-Pk Jitter in Seconds | Degrees | Unit Interval | Unit Interval | Jitter Power |
| 6.43E-09=one cycle | Degrees (Pk-Pk) | Pk-Pk Units | RMS Units | dBui |
| normalized | normalized | normalized | ||
| 1.00E-10 | 5.60 | 0.015552 | 0.0022217 | -53.07 |
| 2.00E-11 | 1.12 | 0.003110 | 0.0004443 | -67.05 |
| (1/7 of Pk-Pk | ||||
| approximation) |
Table 1
Jitter Bandwidth and Spectral Content
The displacement of edges in Fig.1 is a result of noise. Noise has spectral content as well as power.
Consequently, the edge jitter in Fig.1 also has spectral content. The edges in Figure 1 vary randomly
with time, however the noise that causes the jitter is not necessarily uniform over all frequencies.
Jitter due to 10 kHz noise could be greater or less than jitter due to 100 kHz noise. Spectral content
of clock jitter differs greatly depending on the technique used to generate the clock. Measured jitter
also varies with measurement technique and jitter bandwidth. Improperly specified or measured jitter
might result in unnecessary costs, or poor system performance.
See references [2,3] for additional
information on defining and specifying jitter in telecom systems. Jitter characteristics of various
clock sources is discussed later in this article.
How Does Jitter Effect System Performance
The effects of jitter on communication systems are well beyond the scope of this discussion. Refer to
references [1,4] for a more thorough treatment.
A simple discussion may help to understand the deleterious effects of jitter in digital systems.
Every bit of data transmitted over synchronous communication systems is sampled for its value at the
receiver. The sampled data can only have the value of logical one or zero. The optimum point for sampling
data is at the center of each transmit clock cycle. In order to perform this function, the receiver
aligns its own clock with the clock used to transmit the data. Figs.2, a, b, and c represent ideal,
typical, and corrupted datastreams respectively. Commonly referred to as an "eye diagram" each graph is a
cumulative graphical portrait of the edge placement due to noise or jitter. Ideally, sampling occurs at
the center of the "eye". As edge jitter increases, the apparent eye begins to close. As a result,
the likelihood of an error, i.e.
. mistaking a logical one for a zero is more likely. Jitter due to
oscillator noise is only one source of jitter in a telecom system. System designers must consider many
sources of noise in telecom systems. The jitter introduced by clock sources is one component of noise,
and becomes only one part of an "error budget" that must be weighed against performance
requirements and cost.
![]() Fig. 2a |
![]() Fig. 2b |
![]() Fig. 2c |
|
Measurement Techniques
Time Domain Measurements
Edge to Edge Jitter Using a Delay Line
A true measure of clock jitter is the accurate position of clock edges over time. The most direct
method of examining the placement of edges would be to look at the edges using an oscilloscope.
Unfortunately, using standard oscilloscope techniques it is impossible to identify individual clock
edges in absolute time. Any jitter measured with a standard oscilloscope is due to trigger instability.
As a result, direct waveform measurements using an oscilloscope (even a very good oscilloscope) are
not valid measurements of jitter. An additional technique is used to locate the reference edge,
discriminate with time, and examine the jitter on following edges. Figure 3 illustrates this method
with a typical configuration.

Figure 3
The output of the unit under test is fed into splitter/delay line. The non-delayed output of the
splitter is fed to the external trigger input of the oscilloscope (a CSA-803 in this case). The
delayed output of the DL-11 is connected to the input of the oscilloscope. By examining the clock-stream
at a time after the trigger equal to the delay used (in this case, 47 nsec), the trigger-edge is located.
After the triggered edge has been identified, the next edge is examined. A histogram plot is then
produced of the measured jitter of the second edge.
A CSA-803 is used for its statistical and histogram
capabilities. This is a useful technique limited by the length of the delay line and the speed/sensitivity
of the oscilloscope. For all frequencies greater than 1/(2![]()
d), the measurement is limited by the noise of the oscilloscope. Below
1/(2![]()
d), the sensitivity
drops approximately 20 dB/decade. For the 47-nsec delay shown in fig 3, the corner frequency occurs
at 3.3 MHz. All jitter due to frequencies above 3.3 MHz can be resolved to approximately 5 ps using the
CSA-803. Jitter at 330 kHz can not be resolved below 50 ps. In a similar manner, jitter at 33 kHz can
not be measured below 500 ps. Figure 4 is a plot of RMS jitter sensitivity using a 47-nsec delay line.
It is critical to understand the advantages and limitations of this measurement method. For the
numerical example given, low frequency jitter below 300 kHz would not be seen. Conversely, jitter due
to sidebands 3.0 MHz offset or more could easily be identified. This test method is appropriate when
measuring oscillators that employ direct frequency multiplication or where low frequency jitter is not
considered. (See previous description of Pattern Jitter).

Sens(f)=
Where:
Sens(f)=jitter sensitivity
d=amount of delay used
f=the jitter frequency to be measured
r=oscilloscope resolution
Fig. 4
Jitter Measurements Using Phase Locked Loops
It was noted that the length of the delay line limits resolution when measuring edge jitter. In order
to measure jitter below 100 Hz offset, one needs merely to order up about three hundred miles of very
low loss delay line. In lieu of such a device, phase locked loops are used for a variety of noise
measurements.

Fig. 5
Figure 4 shows the basic elements of a phase locked loop (PLL) used to measure the noise of a clock source.
Gardner [5],
Best [6], and
Woolover [7] are three excellent references
for understanding PLLs. Some key loop requirements follow:
Jitter Measurements
Using Phase Locked Loops: Interpreting the Data.
Specifying Jitter Performance
Good jitter performance and low cost are not mutually exclusive as long as:
System Requirements:
Although it is impossible to address all possible variations, some general recommendations based on years
of oscillator manufacturing may be helpful.
While not a complete survey of all applications, Table 1 is a starting point for specifying oscillator
performance. Jitter above 1 kHz is considered high frequency jitter.
Application Performance Requirements
| System Application | Degrees of Difficulty | Low Frequency | High Frequency | Possible |
| Jitter Importance | Jitter Importance | Type | ||
| see Table 2 |
||||
| Radar | Very Difficult Noise Application | Critical | Critical | A, B, C, D |
| Ultrasound/MRI | Very Difficult Noise Application | Critical | Critical | A, B, C, D |
| Navigation/GPS | Difficult Noise Application | Critical | High | A, B, C, D |
| Transmission Systems (telecom) | ||||
| Public Network | Moderate Noise Application | Moderate | Moderate | A, B, C, D |
| Private Network (LAN) | Generally Easiest Application | Low | Low | A, B, F |
| Frequency Synthesis (see note 1) | ||||
| Low Freq. Reference | Moderate-Critical | Moderate-Low | A, B, C | |
| High Frequency Source | Moderate-Low | Moderate-Critical | D, E, F |
Table 1
Note 1: Overall jitter performance is highly dependent on loop parameters.
Clock Generation:
Various methods may be employed to generate high frequency clocks.
Performance may vary significantly based on the technique used. Below 20 MHz, it can be assumed that
direct crystal frequency generation is sufficient for all but the most critical requirements. Low
Noise options should be considered for low jitter applications for 20 MHz and above. Table 2 may be
used as a starting point to select a cost-effective solution. Variations and combinations of methods
listed in table 2 could also be optimal solutions.
Jitter Performance
| Technique | Cost | LF Jitter | HF Jitter | Comments | Type |
| 1 is lowest | 1 is best | 1 is best | |||
| 3 is highest | 3 is worst | 3 is worst | |||
| Direct Clock/TCXO | 1 | 2 | 1 or 2 | Very Good Jitter | A |
| Direct VCXO | 2 | 1 | 1 or 2 | Very Good Jitter | B |
| Direct Oven | 3 | 1 | 1 | Excellent Jitter | C |
| Tuned Multiplication | 2 | 1 | 1 or 2 | Periodic Jitter | D |
| Discrete PLL | 2 | 2 | 2 or 3 | Good Jitter | E |
| Monolithic PLL | 1 | 3 | 2 or 3 | Close in Jitter is Poor | F |
Table 2
CONCLUSION
To correctly specify performance of frequency sources both jitter frequency and amplitude should be
considered. This requires an understanding of jitter, measurement techniques and their limitations.
Time spent to determine system needs will result in fewer problems and less time spent fixing those
problems later on. It will also determine a cost-effective approach for each application.
In this paper, we discussed the definition of jitter, the units used to describe it, and why jitter is
an important parameter. We also reviewed techniques used to measure jitter as well as applications and
typical performance based of various kinds of oscillators. The discussion is by no means complete,
but should give the reader enough information to understand the issues involved. Industry standards
were listed, as well as references for further reading. It is hoped that this paper is useful and
considered a good starting point for understanding and specifying jitter.